Finite element analysis of human lumbar vertebrae in internal fixation system model with different bone density
Julaiti Maitirouzi1, Huiqing Luo1, Zhihao Zhou2
1College of Mechanical Engineering, Xin jiang University, Urumqi, Xin jiang, China.
Purpose:
To evaluate the biomechanics effect of modified cortical bone screw technique (MCBT) with other traditional internal fixation systems on lumbar osteoporotic wet specimen.
Methods:
Four different finite element models were established using CT data: (1) lumbar osteoporosis model without internal fixation system; (2) traditional pedicle screw technology (TT) model; (3) traditional cortical bone screw technology (CBT) model; (4) MCBT model. The changes of global displacement, intervertebral disc displacement of all models and internal fixation system Von Mises stress among the three models were compared under the same physiological load.
Results:
Compared with the other three models, the total displacement of the modified CBT screw model was the smallest, with the maximum displacement of 0.216 mm; The intervertebral disc displacement of the modified CBT screw model was the smallest, with the maximum displacement of 0.149 mm; the internal fixation system Von Mises stress of the modified CBT screw technique model was the largest compared with the other three models, The maximum Von Mises stress is 232.73 MPa.
Conclusion:
Compared to traditional pedicle screw and traditional CBT, MCBT has better mechanical stability, and it is of certain clinical application value.
Related Concept Videos
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...