Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Cycloplegic Autorefraction as a Substitute for Cycloplegic Retinoscopy in the Pediatric Population
Purpose:
To evaluate whether cycloplegic autorefraction can provide similar results as cycloplegic retinoscopy, allowing more comprehensive ophthalmologists to be comfortable in managing pediatric refractive error and refractive amblyopia.
Methods:
This retrospective chart review was performed to determine the mean difference in sphere, cylinder, and axis between cycloplegic autorefraction and retinoscopy, both of which were obtained on the same eye at least 30 minutes after cycloplegia and dilation with a mixed solution of tropicamide, cyclopentolate, and phenylephrine.
Results:
A total of 34 eyes (18 right, 16 left) from 18 patients were included in the analysis. Mean sphere difference between cycloplegic autorefraction and retinoscopy was 0.044 ± 0.278 diopters (D) (95% CI: -1.275 to 1.363 D), mean cylinder difference was -0.081 ± 0.236 D (95% CI: -0.706 to 0.544 D), and mean axis difference was 7.059 ± 19.676 degrees (95% CI: -32.527 to 38.878 degrees). Mean differences in sphere, cylinder, and axis were not statistically significant (P = .362, .0541, and .377, respectively).
Conclusions:
In this small sample population, cycloplegic autorefraction was comparable to cycloplegic retinoscopy. Recognition of amblyopia should still prompt evaluation by a pediatric ophthalmologist. Further research is necessary to confirm whether uncomplicated refractive error in children may be sufficiently detected and managed by a comprehensive ophthalmologist. [.