Related Experiment Video
Updated: Jul 15, 2026

Bacterial Detection & Identification Using Electrochemical Sensors
Published on: April 23, 2013
A V-Shaped Microcantilever Sensor Based on a Gap Method for Real-Time Detection of E. coli Bacteria
1Department of Mechanical and Materials Engineering, Queen's University, Kingston, ON K7L 3N6, Canada.
Abstract:
This paper presents a dynamic-mode microcantilever sensor based on a gap method. The sensor has a V-shaped microcantilever and a fixed structure at a distance of 2 µm from its free end. The microcantilever is excited by applying an ac electric potential (3 Vp) to its piezoelectric pads and vibrates at its fundamental resonant frequency. An independent ac electric potential (200 kHz, 15 Vpp) is applied to the fixed structure. This creates a non-uniform electric field with its maxima at the gap and exerts a dielectrophoresis (DEP) force. The DEP force attracts and adsorbs the E. coli bacteria to the cantilever edge at the gap. The binding of the bacteria to the cantilever creates a shift in the resonant frequency of the microcantilever sensor, which is detected by a laser vibrometer. The real-time detection of E. coli bacteria samples, diluted in distilled water, was performed for concentrations of 105-103 cells/mL and the real-time frequency shifts were -2264.3 to -755 Hz in 4 min, respectively. The tests were expanded to study the effect of the electric potential amplitude (10, 12, 15 Vpp) and higher frequency shifts were observed for higher amplitudes.