Automated Focal Plane Merging From a Stack of Gonioscopic Photographs Using a Focus-Stacking Algorithm.
Masato Matsuo1, Nana Kozuki1, Yuina Inomata1
1Department of Ophthalmology, Shimane University Faculty of Medicine (Matsuo, Kozuki, Inomata, Tanito), Izumo, Japan; NIDEK CO., LTD., Gamagori, Japan (Kumagai, Shiba, Hamaguchi).
Translational Vision Science & Technology
|April 22, 2022
Summary
Automated focus-stacking enhances gonio-photographs, deepening the field of view (DOF) and improving image clarity. This technique aids in the clinical assessment of angle pathology for better glaucoma care.
Area of Science:
- Ophthalmology
- Medical Imaging
- Image Processing
Background:
- Gonio-photography is crucial for assessing angle pathology in glaucoma.
- Limited depth of field (DOF) in traditional gonio-photographs can hinder comprehensive angle assessment.
- Automated focus-stacking offers a potential solution to improve image quality and diagnostic utility.
Purpose of the Study:
- To evaluate the effectiveness of automated focal plane merging using focus-stacking algorithms for gonio-photographs.
- To determine if this technique can enhance the depth of field (DOF) and informativeness of images for diagnosing angle pathology.
Main Methods:
- A cross-sectional study involving 16 eyes from glaucoma patients at Shimane University Hospital.
- Gonio-photographs were captured with varying depths of field (DOF).
- Automated focus-stacking was applied, and resulting images were compared to best-focused images via subjective (masked observers) and objective (Laplacian energy) assessments.
Main Results:
- Automated image processing and focus-stacking were successful in all cases.
- A significant deepening of DOF was achieved in 99.6% of images, and informativeness improved in 84.4% (P < 0.0001).
- Objective assessment showed a significant increase in Laplacian energy (image sharpness) in 94.9% of images (P < 0.0001).
Conclusions:
- Automated focal plane merging effectively increases the DOF of gonio-images compared to standard best-focused images.
- This technique provides subjective and objective improvements, making it valuable for clinical angle pathological assessment.
- The automated focus-stacking algorithm facilitates angle assessment and supports glaucoma care by providing informative, deep-focus images.