Related Experiment Video
Updated: Sep 26, 2025

In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Numerical Assessment of the Risk of Abnormal Endothelialization for Diverter Devices: Clinical Data Driven Numerical
Denis Tikhvinskii1, Julia Kuianova1, Dmitrii Kislitsin2
1Lavrentyev Institute of Hydrodynamics SB RAS, Lavrentiev Avenue 15, 630090 Novosibirsk, Russia.
Abstract:
Numerical modeling is an effective tool for preoperative planning. The present work is devoted to a retrospective analysis of neurosurgical treatments for the occlusion of cerebral aneurysms using flow-diverters and hemodynamic factors affecting stent endothelization. Several different geometric approaches have been considered for virtual flow-diverters deployment. A comparative analysis of hemodynamic parameters as a result of computational modeling has been carried out basing on the four clinical cases: one successful treatment, one with no occlusion and two with in stent stenosis. For the first time, a quantitative assessment of both: the limiting magnitude of shear stresses that are necessary for the occurrence of in stent stenosis (MaxWSS > 1.23) and for conditions in which endothelialization is insufficiently active and occlusion of the cervical part of the aneurysm does not occur (MaxWSS < 1.68)—has been statistacally proven (p < 0.01).