A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Discordant Hemoglobin Values Determined by Blood Gas Analyzer and Hematology Analyzer Synchronously Is a Warning for
1Biochemistry Laboratory, Kepez Public Hospital, Hüsnü Karakaş Mah Güneş Cad No: 124/1, Antalya, Turkey.
A patient may have two contemporary hemoglobin test results; one received from a hematology analyzer and the other from a blood gas analyzer. Such results may pose a dilemma if validated without investigation. A 27-year-old female patient presented to the emergency department because of worsening dyspnea and hemoptysis. Her hemogram revealed Hb 4.5 g/dL; Hct 16.1% while blood gas analysis showed Hb 8.1 g/dL; Hct 25.3%. The results were evaluated to be suspicious. All results were rejected and new requests were made. Hb 9.8 g/dL; Hct 31.9% were seen in repeated hemogram while blood gas analysis showed Hb 10.2 g/dL; Hct 31.5%. It was understood that the nurse used the vein with an intravenous fluid for sampling the first hemogram. A multitude of factors affect the accuracy of hemoglobin concentration measurements. Literature suggests that hemoglobin values of hematology analyzers and blood gas analyzers should be in harmony. Any discordance must be an alarm to stop and review the test results before validation.
A patient may have two contemporary hemoglobin test results; one received from a hematology analyzer and the other from a blood gas analyzer. Such results may pose a dilemma if validated without investigation. A 27-year-old female patient presented to the emergency department because of worsening dyspnea and hemoptysis. Her hemogram revealed Hb 4.5 g/dL; Hct 16.1% while blood gas analysis showed Hb 8.1 g/dL; Hct 25.3%. The results were evaluated to be suspicious. All results were rejected and new requests were made. Hb 9.8 g/dL; Hct 31.9% were seen in repeated hemogram while blood gas analysis showed Hb 10.2 g/dL; Hct 31.5%. It was understood that the nurse used the vein with an intravenous fluid for sampling the first hemogram. A multitude of factors affect the accuracy of hemoglobin concentration measurements. Literature suggests that hemoglobin values of hematology analyzers and blood gas analyzers should be in harmony. Any discordance must be an alarm to stop and review the test results before validation.
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Measurement of Blood Pressure
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Oxygen Transport in the Blood