Related Experiment Video
Updated: Sep 26, 2025

Creating Radio-cephalic Arteriovenous Fistula in the Forearm with a Modified No-Touch Technique
Published on: April 1, 2022
Study on the Safety of the New Radial Artery Hemostasis Device
Baofeng Wu1, Ruixin Zhang2, Chendi Liang2
1Department of Cardiology, First Hospital of Shanxi Medical University, No. 85, JieFang South Road, Yingze District, Taiyuan, Shanxi 030001, China.
Objective:
At present, the use of particular radial hemostatic devices after coronary angiography (CAG) or percutaneous coronary intervention (PCI) has become the primary method of hemostasis. Most control studies are based on the products already on the market, while only a few studies are on the new hemostatic devices. The aim of this study is to compare a new radial artery hemostasis device which is transformed based on the invention patent (Application number: CN201510275446) with TR Band (Terumo Medical) to evaluate its clinical effects.
Methods:
In a prospective randomized clinical trial, 60 patients after CAG or PCI were randomly divided into two groups, patients in the trial group (CD group) using a new radial artery hemostasis device to stop bleeding and the control group (TR group) using the TR Band. The method is to collect relevant data of the two groups and compare the differences in hemostasis, local complications, and patient discomfort between the two groups.
Results:
The hemostatic devices in both groups achieved adequate hemostasis, and there was no failure to stop bleeding. The new radial artery hemostasis device was better than the TR band in pain and swelling (P < 0.05). There were no significant differences in bleeding, hematoma, ecchymosis, skin damage, and local infection between the two groups (P > 0.05).
Conclusions:
The sample of the new radial artery hemostasis device can stop bleeding effectively at the puncture site after CAG or PCI and is not inferior to the TR Band balloon hemostatic device in safety and is better in comfort.
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation