Related Experiment Video
Updated: Sep 25, 2025

Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
Bioimpedance spectroscopy: Is a picture worth a thousand words?
Maha Mohamed1, Jim Matthie2, Stanley L Fan1
1Department of Renal Medicine and Transplantation, Barts Health NHS Trust, London, UK.
Abstract:
Volume status can be difficult to assess in dialysis patients. Peripheral edema, elevated venous pressure, lung crackles, and hypertension are taught as signs of fluid overload, but sensitivity and specificity are poor. Bioimpedance technology has evolved from early single frequency to multifrequency machines which apply spectroscopic analysis (BIS), modeling data to physics-based mixture theory. Bioimpedance plots can aid the evaluation of hydration status and body composition. The challenge remains how to use this information to manage dialysis populations, particularly as interventions to improve over hydration, sarcopenia, and adiposity are not without side effects. It is therefore of no surprise that validation studies for BIS use in peritoneal dialysis patients are limited, and results from clinical trials are inconsistent and conflicting. Despite these limitations, BIS has clinical utility with potential to accurately evaluate small changes in body tissue components. This article explains the information a BIS plot ("picture") can provide and how it can contribute to the overall clinical assessment of a patient. However, it remains the role of the clinician to integrate information and devise treatment strategies to optimize competing patient risks, fluid and nutrition status, effects of high glucose PD fluids on membrane function, and quality of life issues.