Related Experiment Video
Updated: Sep 25, 2025

Robotic D3 Partial Duodenal Resection with Primary Side-to-Side Anastomosis
Published on: December 15, 2023
Clinical Adverse Events and Device Failures Reported for the Captivator and Duette Endoscopic Mucosal Resection (EMR)
Lauren D Hagenstein1, Katsiaryna Dubrouskaya1, Daryl Ramai2
1Rocky Vista University College of Osteopathic Medicine, Ivins.
Introduction:
Endoscopic mucosal resection (EMR) is a widely used technique for the removal of precancerous or select cancerous mucosal-based lesions. Two systems used for EMR of upper gastrointestinal lesions are the Duette EMR kit and the Captivator EMR kit. The aim of this study is to analyze the reports submitted to the Food and Drug Administration's (FDA) Manufacturer and User Facility Device Experience (MAUDE) database for patient-related adverse events and device failures associated with these EMR kits.
Methods:
We analyzed postmarketing data from the FDA MAUDE from August 2004 to August 2021.
Results:
Eighty-four MAUDE reports were found in the study period from August 2004 to August 2021. In total, device issues comprised 36.9% of the reports (n=31) and patient adverse events comprised 63.09% of the reports (n=53). Device issues that were reported included, but were not limited to, band placement/deployment failure (8.1%), suture breakage (7.31%), and snare breakage (4.94%). Patient adverse events that were reported included, but were not limited to, bleeding (25.9%), perforation (7.31%), and snare breakage (4.94%).
Conclusion:
Captivator EMR and Duette EMR have similar patterns of technical failure and clinical adverse events. More MAUDE reports exist for Duette EMR, possibly due to its longer duration of commercial availability.
Related Concept Videos
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Endoscopic Procedures III: Video Capsule Endoscopy
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: