A comprehensive comparison between ISAC and ALEX2 multiplex test systems
Anouk C M Platteel1, Pieter van der Pol1, Jean-Luc Murk1
1Microvida Laboratory of Medical Microbiology and Immunology, Elisabeth-TweeSteden Ziekenhuis, Tilburg, Noord-Brabant, Netherlands.
Clinical Chemistry and Laboratory Medicine
|April 26, 2022
Summary
The ALEX 2 multiplex assay shows high agreement with the ImmunoCAP ISAC system for diagnosing type I hypersensitivity. This allergy testing platform offers reliable results for specific IgE antibody detection.
Area of Science:
- Allergy and Immunology
- Diagnostic Testing
- Molecular Diagnostics
Background:
- Type I hypersensitivity diagnosis relies on clinical history, provocation tests, and immunological assays.
- Multiplex specific IgE (sIgE) testing allows simultaneous detection of sIgE antibodies against numerous allergens.
- Evaluating novel diagnostic platforms is crucial for improving allergy diagnostics.
Purpose of the Study:
- To assess the performance of the ALEX 2 multiplex platform for allergy diagnostics.
- To compare the ALEX 2 test results with the established ImmunoCAP ISAC system.
- To determine the concordance and agreement metrics between the two sIgE testing platforms.
Main Methods:
- Serum samples from 49 patients with known ISAC results were analyzed using the ALEX 2 platform.
- Comparison involved 103 common allergen components present on both platforms.
- Statistical analysis included Cohen's kappa, negative percent agreement (NPA), and positive percent agreement (PPA).
Main Results:
- Overall agreement between ALEX 2 and ISAC was high at 94%.
- Negative percent agreement (NPA) was 95%, and positive percent agreement (PPA) was 90%.
- Kappa values ranged from 0.60 to 0.92, indicating moderate to almost perfect agreement; 59% of discrepancies involved weak positive results.
Conclusions:
- The ALEX 2 multiplex assay demonstrates high concordance with the ImmunoCAP ISAC system for common allergen components.
- The findings support the reliability of the ALEX 2 platform for sIgE testing in allergy diagnostics.
- Further investigation into discrepancies, particularly weak positive results, may refine clinical interpretation.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
315
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
315
Multiple Comparison Tests
4.0K
Multiple comparison test, abbreviated as MCT, is a post hoc analysis generally performed after comparing multiple samples with one or more tests. An MCT will help identify a significantly different sample among multiple samples or a factor among multiple factors.
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
4.0K