Background:
Severe COVID-19 can necessitate multiple organ support including veno-venous extracorporeal membrane oxygenation (vvECMO) and renal replacement therapy. The therapy can be complicated by venous thromboembolism due to COVID-19-related hypercoagulability, thus restricting vascular access beyond the vvECMO cannula. Although continuous renal replacement therapy can be performed via a vvECMO circuit, studies addressing sustained low-efficiency dialysis (SLED) integration into vvECMO circuits are scarce. Here we address the lack of evidence by evaluating feasibility of SLED integration into vvECMO circuits.
Methods:
Retrospective cohort study on nine critically ill COVID-19 patients, treated with integrated ECMO-SLED on a single intensive care unit at a tertiary healthcare facility between December 2020 and November 2021. The SLED circuits were established between the accessory arterial oxygenator outlets of a double-oxygenator vvECMO setup. Data on filter survival, quality of dialysis, and volume management were collected and compared with an internal control group receiving single SLED.
Results:
This study demonstrates general feasibility of SLED integration into existing vvECMO circuits. Filter lifespans of ECMO-SLED compared with single SLED are significantly prolonged (median 18.3 h vs. 10.3 h, p < 0.01). ECMO-SLED treatment is furthermore able to sufficiently normalize creatinine, blood urea nitrogen, and serum sodium, and allows for adequate ultrafiltration rates.
Conclusions:
We can show that ECMO-SLED is practical, safe, results in adequate dialysis quality and enables sufficient electrolyte and volume management. Our data indicate that SLED devices can serve as potential alternative to continuous-veno-venous-hemodialysis for integration in vvECMO circuits.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis I: Introduction
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Continuous Renal Replacement Therapy