Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
Rapid cyclic ion mobility separations of monosaccharide building blocks as a first step toward a high-throughput
1Department of Chemistry, University of Utah 315 South 1400 East, Room 2020 Salt Lake City Utah 84112 USA gabe.nagy@utah.edu.
Abstract:
Herein we present a new high-throughput screening method for carbohydrate syntheses based on cyclic ion mobility spectrometry-mass spectrometry (cIMS-MS)-based separations. We rapidly resolved the α/β anomers for carbohydrates with varying protecting groups after only 5 m of cIMS-MS separation and also detected their respective unwanted anomeric impurities at levels lower than 2%. All experiments were performed in 1 minute of total acquisition time demonstrating our method's high-throughput nature. Our methodology was also extended to the separation of an isomeric mixtures of two protected disaccharides illustrating its utility beyond only monosaccharides. We envision our presented workflow as a first step toward the development of a high-throughput screening platform for the rapid and sensitive detection of α/β anomeric selectivities and for trace isomeric/isobaric impurities.
More Related Videos
12:02Automated Modular High Throughput Exopolysaccharide Screening Platform Coupled with Highly Sensitive Carbohydrate Fingerprint Analysis
Published on: April 11, 2016
07:12An Efficient Sample Preparation Method to Enhance Carbohydrate Ion Signals in Matrix-assisted Laser Desorption/Ionization Mass Spectrometry
Published on: July 29, 2018
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...