Active Damping of Air-backed Ultrasonic Transducers Using Arbitrary Waveform Generators.
1Biomedical Engineering, University of Southern California Los Angeles, USA.
This study demonstrates active damping to enhance narrowband ultrasound transducers. This technique broadens transducer bandwidth, improving performance for advanced ultrasound applications.
Area of Science:
- Acoustics and Ultrasound Engineering
- Biomedical Engineering
- Transducer Technology
Background:
- Advanced ultrasound technologies like high-intensity focused ultrasound and acoustic radiation force imaging demand sophisticated transducer designs.
- These designs typically require transducers with wide power ranges, high sensitivity, and broad bandwidth capabilities.
- A key challenge is achieving broadband performance from a single transducer suitable for diverse applications.
Purpose of the Study:
- To demonstrate the feasibility of achieving broadband capability in air-backed, narrowband ultrasound transducers.
- To investigate the use of active damping as a method to broaden transducer bandwidth.
Main Methods:
- Active damping was implemented using arbitrary waveform generators to cancel unwanted oscillations.
- A modified 1-D KLM model in Matlab was developed to guide waveform design and optimization.
- Waveform optimization employed a minimization function to reduce signal ripple.
Main Results:
- In simulations, active damping increased the -3 dB transmitted bandwidth from 10% to 44% for 1.5-cycle excitation.
- For 1-cycle excitation, the bandwidth increased from 11.4% to 63.8% in the model.
- Experimental results showed comparable increases in bandwidth, validating the simulation findings.
Conclusions:
- Active damping is a feasible technique for enhancing the bandwidth of narrowband ultrasound transducers.
- This method offers a pathway to achieving broadband capabilities in existing transducer designs.
- The findings support the potential for using a single transducer for multiple advanced ultrasound applications.
More Related Videos
08:32Assembly and Characterization of an External Driver for the Generation of Sub-Kilohertz Oscillatory Flow in Microchannels
Published on: January 28, 2022
11:44Real-Time DC-dynamic Biasing Method for Switching Time Improvement in Severely Underdamped Fringing-field Electrostatic MEMS Actuators
Published on: August 15, 2014
Related Concept Videos
Types of Damping
Damped Oscillations
Although friction and other non-conservative...
Sound Waves: Interference
Design Example: Underdamped Parallel RLC Circuit
Starting with a fixed...
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Concept of Resonance and its Characteristics