Dynamic Monitoring of Seroconversion using a Multianalyte Immunobead Assay for Covid-19
Published on: February 16, 2022
Validation and performance of a multiplex serology assay to quantify antibody responses following SARS-CoV-2
Deidre Wilkins1, Anastasia A Aksyuk1, Alexey Ruzin1
1Translational Medicine, Vaccines and Immune Therapies BioPharmaceuticals Medical AstraZeneca Gaithersburg MD USA.
Objectives:
Robust, quantitative serology assays are required to accurately measure antibody levels following vaccination and natural infection. We present validation of a quantitative, multiplex, SARS-CoV-2, electrochemiluminescent (ECL) serology assay; show correlation with two established SARS-CoV-2 immunoassays; and present calibration results for two SARS-CoV-2 reference standards.
Methods:
Precision, dilutional linearity, ruggedness, analytical sensitivity and specificity were evaluated. Clinical sensitivity and specificity were assessed using serum from prepandemic and SARS-CoV-2 polymerase chain reaction (PCR)-positive patient samples. Assay concordance to the established Roche Elecsys® Anti-SARS-CoV-2 immunoassay and a live-virus microneutralisation (MN) assay was evaluated.
Results:
Standard curves demonstrated the assay can quantify SARS-CoV-2 antibody levels over a broad range. Assay precision (10.2-15.1% variability), dilutional linearity (≤ 1.16-fold bias per 10-fold increase in dilution), ruggedness (0.89-1.18 overall fold difference), relative accuracy (107-118%) and robust selectivity (102-104%) were demonstrated. Analytical sensitivity was 7, 13 and 7 arbitrary units mL-1 for SARS-CoV-2 spike (S), receptor-binding domain (RBD) and nucleocapsid (N) antigens, respectively. For all antigens, analytical specificity was > 90% and clinical specificity was 99.0%. Clinical sensitivities for S, RBD and N antigens were 100%, 98.8% and 84.9%, respectively. Comparison with the Elecsys® immunoassay showed ≥ 87.7% agreement and linear correlation (Pearson r of 0.85, P < 0.0001) relative to the MN assay. Conversion factors for the WHO International Standard and Meso Scale Discovery® Reference Standard are presented.
Conclusions:
The multiplex SARS-CoV-2 ECL serology assay is suitable for efficient, reproducible measurement of antibodies to SARS-CoV-2 antigens in human sera, supporting its use in clinical trials and sero-epidemiology studies.