Background:
The first surgeries with CI532 showed an effect of the proximity of the electrode to the modiolus on the Evoked Compound Action Potentials (ECAPs).
Objectives:
Objectives of the study were to investigate the effect of the "pullback" procedure on intraoperative ECAP responses in three different electrode array positions and additionally to compare behavioral thresholds with the thresholds obtained in a group of patients using the standard insertion. The hypothesis of this study is that pullback will cause lower ECAPs and behavioral thresholds.
Patients:
The study included 40 patients, 20 in the pullback insertion group and 20 in the standard insertion group (without pullback).
Method:
During insertion of the CI532 electrode array, ECAP was performed in three different positions for the pullback group: at initial insertion, at over-insertion, and after pullback. Insertion was monitored by fluoroscopy. In the standard group, ECAP was performed at the initial position, which is also the final position. ECAP thresholds (T-ECAPs) were compared within subjects at the initial and the final position in the pullback group and between groups in the final positions of the pullback and standard groups. Programming levels (C- and T-levels) were compared between the two groups 1 year after switch-on.
Results:
Intraoperative measurements pullback shows lower average T-ECAPs after pullback compared to thresholds in initial position. Comparison of intraoperative T-ECAPs at the final positions showed no statistically significant difference between the pullback group and the standard insertion group. Furthermore, 1 year after switch-on there was no statistically significant difference in C- and T-levels between the two groups.
Conclusion:
The pullback maneuver of the CI532 electrode array after an over-insertion gave significantly lower T-ECAPs compared to the thresholds at the initial position. However, the between-groups analysis of pullback and standard insertion showed neither significantly different T-ECAPs nor different programming levels. Because T-ECAPs and programming levels vary considerably between subjects, large groups are required to detect differences between groups. Additionally, the effect pullback technique to preserving the residual hearing is not known yet.
More Related Videos
09:06Cochlear Implant Surgery and Electrically-evoked Auditory Brainstem Response Recordings in C57BL/6 Mice
Published on: January 9, 2019
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
The Cochlea
Hair Cells
Propagation of Action Potentials
Neurons (nerve cells) have a resting membrane potential, with a slightly negative charge inside compared to outside. This is maintained by ion channels, such as sodium (Na+) and potassium (K+) channels, which control the flow of ions. When a stimulus, like a touch or a signal from another neuron, triggers the neuron, sodium channels open, allowing sodium ions to...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...