Dental imaging using an ultra-high resolution photon-counting CT system
Maurice Ruetters1, Sinan Sen2, Holger Gehrig3
1Section of Periodontology, Department of Operative Dentistry, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. maurice.ruetters@med.uni-heidelberg.de.
Abstract:
Clinical photon-counting CT (PCCT) offers a spatial resolution of about 200 µm and might allow for acquisitions close to conventional dental CBCTs. In this study, the capabilities of this new system in comparison to dental CBCTs shall be evaluated. All 8 apical osteolysis identified in CBCT were identified by both readers in all three PCCT scan protocols. Mean visibility scores showed statistical significant differences for root canals(p = 0.0001), periodontal space(p = 0.0090), cortical(p = 0.0003) and spongious bone(p = 0.0293) in favor of high and medium dose PCCT acquisitions. Overall, both devices showed excellent image quality of all structures assessed. Interrater-agreement showed high values for all protocols in all structures. Bland-Altman plots revealed a high concordance of both modalities with the reference measurements. In vitro, ultra-high resolution PCCT can reliably identify different diagnostic entities and structures relevant for dental diagnostics similar to conventional dental CBCT with similar radiation dose. Acquisitions of five cadaveric heads were performed in an experimental CT-system containing an ultra-high resolution PC detector (0.25 mm pixel size in isocenter) as well as in a dental CBCT scanner. Acquisitions were performed using dose levels of 8.5 mGy, 38.0 mGy and 66.5 mGy (CTDI16cm) in case of PCCT and of 8.94 mGy (CTDI16cm) in case of CBCT. The quality of delineation of hard tissues, root-canals, periodontal-space as well as apical osteolysis was assessed by two readers. Mean visibility scores and interrater-agreement (overall agreement (%)) were calculated. Vertical bone loss (bl) and thickness (bt) of the buccal bone lamina of 15 lower incisors were measured and compared to reference measurements by ore microscopy and clinical probing.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...