Related Experiment Video
Updated: Sep 24, 2025

A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
Noise exposure during robot-assisted total knee arthroplasty
Tim Hönecke1,2, Michael Schwarze3, Matthias Wangenheim4
1Department of Orthopedic Surgery, Hannover Medical School, Anna-von-Borries-Str. 1-7, 30625, Hannover, Germany. tim.hoenecke@diakovere.de.
Noise exposure during robot-assisted total knee arthroplasty (TKA) exceeds occupational safety guidelines. This poses a significant risk for developing noise-induced hearing loss (NIHL) in surgical staff.
Area of Science:
- Orthopedics
- Occupational Health
- Biomedical Engineering
Background:
- Noise exposure during manual total knee arthroplasty (TKA) surpasses occupational guidelines.
- Prolonged exposure to high noise levels increases the risk of noise-induced hearing loss (NIHL) in surgical staff.
- Robot-assisted TKA is increasingly adopted, necessitating an evaluation of its associated noise risks.
Purpose of the Study:
- To assess and compare operating theater noise levels during TKA procedures using three distinct robot systems: MAKO, NAVIO, and CORI.
- To determine if noise exposure from robot-assisted TKA exceeds established occupational safety limits.
- To identify potential risks of NIHL for surgical staff involved in these procedures.
Main Methods:
- Measurements of average and peak sound pressure levels were conducted during TKA procedures utilizing MAKO, NAVIO, and CORI robot systems.
- A class 1 sound level meter was employed for accurate noise data collection.
- Data were compared against recommended guidelines from the National Institute for Occupational Safety and Health (NIOSH).
Main Results:
- All three robot systems (MAKO, NAVIO, CORI) generated noise levels exceeding NIOSH recommended guidelines.
- The MAKO robot exhibited the highest average sound level (93.18 dB(A)).
- The NAVIO robot recorded the highest peak sound level (134.48 dB(C)).
Conclusions:
- Robot-assisted TKA, similar to manual TKA, presents a risk factor for noise-induced hearing loss (NIHL).
- Operating theater staff performing robot-assisted TKA are exposed to noise levels that warrant concern for hearing health.
- Further research is crucial to develop strategies for reducing noise exposure during TKA to protect surgical staff from hearing damage.