Related Experiment Video
Updated: Jun 21, 2026

12:50
Colorimetric Paper-based Detection of Escherichia coli, Salmonella spp., and Listeria monocytogenes from Large Volumes of Agricultural Water
Published on: June 9, 2014
14.7K
Quantitative determination of leukocyte esterase with a paper-based device
Mei-Lin Ho1, Wei-Fang Liu1, Hsin-Yi Tseng1
1Department of Chemistry, Soochow University Taipei 111 Taiwan meilin_ho@gm.scu.edu.tw +886 2 2881 1053 +886 2 2881 9471 ext. 6827.
RSC Advances
|May 6, 2022
Summary
A novel paper-based device accurately quantifies urinary leukocyte esterase (LE) for early urinary tract infection (UTI) detection. This LE-device offers improved sensitivity and specificity compared to standard urine dipsticks for reliable UTI diagnosis.
Area of Science:
- Analytical Chemistry
- Biomedical Engineering
- Point-of-Care Diagnostics
Background:
- Commercially available urine dipsticks for urinary tract infection (UTI) detection have limitations in accurately quantifying leukocyte esterase (LE).
- The quantitative determination of urinary LE for predicting UTI remains uncertain, necessitating improved diagnostic tools.
Purpose of the Study:
- To develop and validate a paper-based analytical device for the quantitative point-of-care detection of LE (LE-PAD).
- To assess the diagnostic performance of the LE-PAD for early UTI detection in inpatients.
Main Methods:
- Fabrication of the LE-PAD using a coating of mixed 3-(N-tosyl-L-alaninyloxy)-5-phenylpyrrole (PE) and 1-diazo-2-naphthol-4-sulfonic acid (DAS) on a silver conducting film.
- Quantitative analysis of LE in urine samples from 21 inpatients with urinary catheters by measuring changes in resistivity.
- Establishment of a quantitative calibration curve and evaluation of diagnostic accuracy using receiver operating characteristic (ROC) analysis.
Main Results:
- The LE-PAD demonstrated a strong predictive value for UTI diagnosis, with an area under the ROC curve of 0.875.
- At an optimal cut-off, the LE-PAD achieved a sensitivity of 87.5% and specificity of 92.3% for UTI diagnosis.
- Compared to standard urine dipsticks (sensitivity 62.5%, specificity 76.9%), the LE-PAD showed significantly higher diagnostic accuracy.
Conclusions:
- The developed LE-PAD is a reliable and quantitative point-of-care test for detecting urinary LE.
- The LE-PAD offers superior diagnostic performance for UTI compared to conventional urine dipsticks.
- This novel device holds promise for improving early UTI detection and management.