Related Experiment Video
Updated: Sep 24, 2025

Biosensor for Detection of Antibiotic Resistant Staphylococcus Bacteria
Published on: May 8, 2013
Quantification of Staphylococcus aureus using surface acoustic wave sensors
Zhangliang Xu1,2, Yong J Yuan1
1Laboratory of Biosensing and MicroMechatronics, School of Materials Science and Engineering, Southwest Jiaotong University Chengdu Sichuan 610031 China yongyuan@swjtu.edu.cn.
Abstract:
Quartz crystal microbalance (QCM), surface acoustic wave (SAW)-Rayleigh and ZnO based SAW-Love sensors were fabricated and their sensitivity was comparatively analyzed for the quantification of Staphylococcus aureus (S. aureus). The SAW based sensors showed response magnitudes up to three times greater than that of the QCM sensor for the same mass loading of S. aureus. The ZnO nanoparticle-based SAW-Love sensor has a maximum mass loading sensitivity of 328.75 Hz ng-1. The SAW-Love sensor achieved a lower limit of detection of 2 × 103 CFU mL-1 compared to the QCM counterpart (2 × 105 to 2 × 106 CFU mL-1) under the same conditions. The SAW-Love sensor could be used as a disposable chip in micro or ultra-trace accurate diagnosis methods.