Mitigating analyte to stable isotope labelled internal standard cross-signal contribution in quantitative liquid
Mirjana Radovanovic1,2, Graham Jones1,2, Richard O Day1,3
1St Vincent's Clinical School, University of New South Wales, Darlinghurst, NSW, Australia.
Background:
Utilising stable isotope labelled internal standards (SIL-IS) in quantitative LC-MS/MS drug analysis is the most widely used approach to normalise for variability during sample quantification processes. However, compounds containing atoms such as Sulphur, Chlorine or Bromine, could potentially cause cross-signal contribution to the SIL-IS from the naturally occurring isotopes, resulting in non-linear calibration curves. A simple, novel method of mitigating the effect is presented here. It entails monitoring of a less abundant SIL-IS isotope, as the precursor ion, of a mass that has no/minimal isotopic contribution from the analyte isotopes.
Methods:
Experiments were conducted on two LC-MS/MS analysers: Waters Xevo TQ-S and Shimadzu 8050. Flucloxacillin (FLX) was used as an example. Two transitions were selected for FLX (m/z 454 → 160 → 295) and one for each of the SIL-IS isotopes (m/z 458 → 160 for the isotope 457 g/mol and m/z 460 → 160 for the isotope 459 g/mol). Assay biases were assessed at three SIL-IS concentrations: 0.7, 7 and 14 mg/L for each isotope.
Results:
When using the SIL-IS isotope m/z 458 → 160 at a concentration of 0.7 mg/L, biases were up to 36.9 % on both instruments. Increasing the SIL-IS concentration to 14 mg/L, reduced the bias to 5.8 %. Using the less abundant isotope, m/z 460 → 160, resulted in biases of 13.9 % at an SIL-IS concentration of 0.7 mg/L.
Conclusions:
Applying this method will mitigate cross-signal contribution from the analyte isotopes to the corresponding SIL-IS, minimise the use of SIL-IS, and, thereby, reduce overall cost.
More Related Videos
08:06Absolute Quantification of Cell-Free Protein Synthesis Metabolism by Reversed-Phase Liquid Chromatography-Mass Spectrometry
Published on: October 25, 2019
09:24Automated Sample Multiplexing by using Combined Precursor Isotopic Labeling and Isobaric Tagging cPILOT
Published on: December 18, 2020
Related Concept Videos
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Isotope Effect
High-Performance Liquid Chromatography: Types of Detectors