Related Experiment Video
Updated: Sep 24, 2025

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Ultra-low-dose intraoperative X-ray imager for minimally invasive surgery: a pilot imaging study
Haewook Park1,2, Kook Nam Han3, Byeong Hyeon Choi4,5
1Department of Biomedical Sciences, Seoul National University College of Medicine, Seoul, Republic of Korea.
Background:
With advances in surgical technology, thoracic surgeons have widely adopted minimally invasive limited-resection techniques to preserve normal tissues. However, it remains difficult to achieve in situ localization of invisible pulmonary nodules during surgery. Therefore, we proposed an in situ ultra-low-dose X-ray imaging device for intraoperative pulmonary nodule localization during minimally invasive surgeries.
Methods:
The proposed device features a hand-held type and consists of a carbon nanotube-based X-ray source and an intraoral dental sensor. In a preclinical study, we created pseudo pulmonary nodules using ex vivo pig lungs. Subsequently, its clinical feasibility was evaluated using ex vivo lung cancer specimens from patients with cancer who had undergone minimally invasive surgery.
Results:
Using the proposed device, we successfully differentiated normal and abnormal tissues from X-ray images of resected lung specimens. In addition, our proposed device only yielded an average radiation dose of 90.9 nGy for a single acquisition of X-ray images and demonstrated excellent temperature stability under consecutive X-ray irradiations. The radiation exposure of our proposed device (0.1±0.0006 µSv/h) was significantly lower than that of conventional C-arm fluoroscopy (41.5±51.8 µSv/h). In both preclinical and clinical studies, the margin of nodule shadows was clearly visualized using the proposed device.
Conclusions:
The proposed device substantially reduced radiation exposure to staff and patients and may allow in situ localization of pulmonary nodules. Our proposed device clearly revealed the margins of lung nodules with radiocontrast injection and showed the potential to identify solid nodules without the use of radiocontrast agents.