Introduction:
The locking attachment plate (LAP) can be added to a locking compression plate (LCP) to allow the fixation of locking screws bicortically around a femoral implant. We aimed to examine surgical and fracture characteristics associated with healing for periprosthetic femur fractures (PPFFx) treated with constructs employing LAP fixation. We hypothesize that the addition of an LAP provides stable peri-implant fixation.
Materials &Methods:
We retrospectively reviewed a consecutive series of 28 PPFFx surgically treated with LCP-LAP constructs by 4 surgeons from 2015-2020. Fractures were classified and grouped using the Vancouver Classification System and included 12 B1, 2 B2, 11 C fractures, and 3 fractures around other stemmed implants. Primary outcome measures included hardware failure such as screw pullout, broken screws, and plate fracture. Clinical complications including infection, non-union, malunion, and reoperation were recorded.
Results:
No LAP failures, screw pullout, or broken screws were observed. Two fractured plates (7.1%) occurred in patients with Vancouver C fracture types. Overall complication rate was 17.9% and included 3 non-unions, 1 deep infection, and 1 implant loosening with painful hardware, each requiring reoperation. Differences were observed between unions and nonunions for total number of screws (12.4 vs 14.7, P = .005) and number of locking screws used (8.04 vs 11.3, P = .03).
Conclusion:
The LAP provides adequate fixation and low failure rates where fixation is required around a well-fixed stem. When failures occur, it is from plate breakage and not due to failure of fixation at the area of plate-stem overlap.
More Related Videos
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...