Related Experiment Video
Updated: Jul 23, 2026

Automation of a Positron-emission Tomography PET Radiotracer Synthesis Protocol for Clinical Production
Published on: October 26, 2018
[68Ga]/[90Y]FAPI-46: Automated production and analytical validation of a theranostic pair
M Nader1, D F Valla2, C Vriamont3
1Klinik für Nuklearmedizin, Universitätsklinikum Essen (AöR), Hufelandstraße 55, D-45147 Essen, Germany.
Introduction:
With respect to the broad application of FAPI-46 in therapy and diagnostics, there is a need for an efficient as well as convenient way for routine production and quality control of the theranostic pair [90Y]Y/[68Ga]Ga-FAPI-46, since no monograph is currently available for radiolabelled FAPI derivatives. The aim of the current work is to create a GMP compliant theranostic set up for the production and quality control of the diagnostic [68Ga]Ga-FAPI-46 as well as the therapeutic drug [90 Y]Y-FAPI-46, which can be the basis for future monographic standards.
Methods:
Sterile [90Y]yttrium chloride solution and a pharmaceutical grade 68Ge/68Ga generator were applied for the labelling of FAPI-46 using the cassette based synthesis module Trasis EASYONE. All chemicals were GMP-grade and excipients were with marketing authorisation. The quality control included test procedures according to Ph. Eur.
Results:
Fully automated synthesis of the theranostic pair [90Y]Y/[68Ga]Ga-FAPI-46 was achieved on the Trasis EasyOne synthesizer with a radiochemical yield of 88 ± 7% and 56 ± 5% with a radiochemical purity of >99%. Stability experiments showed a durability for [68Ga]Ga-FAPI-46 within 4 h and for [90Y]Y-FAPI-46 within 24 h. All obtained specifications and validations were compliant with the European Pharmacopoeia and regulatory guidelines. Both products were successfully applied in cancer patients.
Conclusion:
In the present work, efficient and robust procedures for the automated production and quality control of the theranostic pair [68Ga]/[90Y]FAPI 46 were developed and validated using the same synthetic platform. The described methods were evaluated in accordance with existing guidelines and toxicological limits, which can be a valuable basis for future monographic standards.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
![An Automated Radiosynthesis of [68Ga]Ga-FAPI-46 for Routine Clinical Use](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F66708.jpg&w=3840&q=50)
![Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F68356.jpg&w=3840&q=50)