Magnetic field interactions between current consumer electronics and cardiac implantable electronic devices
Kathryn Xu1, Jay Sengupta2, Susan Casey2
1Heart Rhythm Science Center, Minneapolis Heart Institute and Foundation, 800 E 28th St, Minneapolis, MN, 55407, USA. kathrynxu07@gmail.com.
New iPhones and other electronics with magnets can affect cardiac implantable electronic devices (CIEDs). However, interaction with CIEDs is unlikely beyond 1.5 cm, ensuring safety during daily activities.
Area of Science:
- Biomedical Engineering
- Electromagnetism
- Cardiology
Background:
- Modern electronic devices like the iPhone 12, Apple Watch Series 6, and AirPods utilize magnets for wireless charging.
- These permanent magnets pose a potential risk to cardiac implantable electronic devices (CIEDs), potentially causing asynchronous pacing in pacemakers and suspending arrhythmia detection in defibrillators.
- This research aims to evaluate the impact of static magnetic fields from consumer electronics on CIEDs and inform FDA recommendations.
Purpose of the Study:
- To determine if CIEDs are affected by static magnetic fields from common portable electronics (PE) at any distance.
- To assess the interaction range between PE magnets and CIEDs.
- To provide evidence-based recommendations for consumer safety regarding PE and CIEDs.
Main Methods:
- Magnetic field strength of PE was measured using a Gauss meter.
- Ex vivo testing was performed using a body torso model with CIEDs from Boston Scientific and Medtronic immersed in saline.
- PE was placed at the CIED surface and at distances of 0.5, 1.0, and 1.5 cm to assess magnet mode activation.
Main Results:
- The iPhone 12 exhibited a significantly stronger static magnetic field at its surface compared to the iPhone XR.
- All tested CIEDs exhibited magnet mode activation when PE was placed at the surface of the torso model.
- The maximum distance at which interaction (magnet mode activation) occurred between the tested Apple products and CIEDs was 1.5 cm.
Conclusions:
- The iPhone 12 generates a more potent static magnetic field than its predecessors.
- Portable electronics containing magnets are unlikely to interact with CIEDs beyond a distance of 1.5 cm.
- Given the limited interaction distance, magnet mode activation is improbable during typical daily activities, suggesting consumer safety.