A suspected case of falsely low digoxin and vancomycin concentrations caused by free kappa light chains with PETINIA

Nga Yeung Tang1,2, Kelly Walewski1, Robin Carey-Ballough1

  • 1Department of Pathology and Laboratory Medicine, Beaumont Health, Royal Oak, MI, USA.

Insights

Free kappa light chains can cause falsely low measurements of vancomycin and digoxin levels. This interference in immunoassays necessitates careful interpretation of drug concentrations and clinical correlation.

Area of Science:

  • Clinical Chemistry
  • Immunochemistry
  • Laboratory Medicine

Background:

  • Vancomycin and digoxin require therapeutic drug monitoring due to potential toxicity.
  • Paraprotein interference with vancomycin assays is known, but not with digoxin assays.

Observation:

  • A patient on vancomycin and digoxin showed persistently low serum concentrations by Abbott PETINIA method.
  • Alternative methods revealed toxic drug levels, prompting drug cessation.

Findings:

  • Free kappa light chains, potentially in polymeric form, are suspected to cause falsely low vancomycin and digoxin levels.
  • Turbidity and agglutination in patient samples suggest assay interference.
  • Protein-free filtration partially corrected the drug measurements.

Implications:

  • Clinicians and pharmacists must consider assay interference when drug levels conflict with clinical presentation.
  • Laboratory investigation for interfering substances is crucial for accurate therapeutic drug monitoring.
  • This case highlights a novel interference mechanism in particle-enhanced turbidimetric immunoassay methods.