Correlation of Cardiac Output by Arterial Contour-Derived Cardiac Output Monitor Versus Pulmonary Artery Catheter in
Kusuma Halemani1, Lakshmi Kumar1, Bhadrinath Narayanan1
1Department of Anaesthesia, Amrita Institute of Medical Sciences, Kochi, Kerala, India.
Objective:
Arterial pulse-derived cardiac output monitors are routinely employed to guide hemodynamic management during liver transplant surgery. In this study, we sought to assess the reliability by evaluating the agreement of the cardiac output measured by the FloTrac Vigileo versus pulmonary artery catheter (continuous cardiac output) at specified times during liver transplant.
Methods:
Liver transplant database with cardiac output values measured by FloTrac Vigileo and continuous cardiac output was analyzed retrospectively at a tertiary care hospital. Data were compared at T0: baseline, T1: 1 hour in dissection phase, T2: anhepatic phase, T3: portosystemic shunt, T4: reperfusion, T5: 1 hour after reperfusion, and T6: skin closure. Statistical analysis was done using Bland-Altman analysis and percentage error (<30%) to assess the agreement between cardiac output measured by 2 techniques, Lin's concordance correlation coefficient for quantifying the agreement and 4-quadrant plots to compare the trends of cardiac output.
Results:
Bland-Altman analysis showed mean cardiac output ± standard deviation L min-1 (95% CI) at T0: 0.2 ± 2.09 (-3.9 to 4.3), T1: 0.53 ± 3.0 (-5.4 to 6.4), T2: 0.47 ± 2.1(-3.7 to 4.6), T3: 0.31 ± 1.9 (-3.4 to 4.0), T4: 0.44 ± 2.15 (-3.8 to 4.7), T 5:0.69 ± 1.9. (-2.9 to 4.3), and at T6: 0.43 ± 2.25 (-4.0 to 4.8). Percentage error was 44%-72% and concordance correlation coefficient was poor (<0.65) at all points.
Conclusions:
There is poor agreement between the cardiac output measured by FloTrac and pulmonary artery catheter among liver transplant recipients. The need for superior hemodynamic monitoring is mandated in liver transplant.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Cardiac Catheterization II: Right Heart Catheterization