The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Application of a novel osteotomy instrumentation as a substitute tool in total hip arthroplasty
Yuehao Hu1, Jingwei Zhang1, Ziyang Sun2
1Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200011, People's Republic of China.
Background:
Mechanical failure, power shortage, and inadvertent contamination of the oscillating saw occasionally occurs in actualizing femoral neck osteotomy during total hip arthroplasty (THA); however, no appropriate alternative solution is currently available. This study aimed to introduce a novel osteotomy instrumentation (fretsaw, jig, cable passer hook) as a substitute tool while the oscillating saw was unavailable during THA.
Methods:
This study included 40 patients (40 hips) who underwent femoral neck osteotomy during primary THA using the new osteotomy instrumentation (n = 20) and the oscillating saw (n = 20). Clinical data and intraoperative findings of all patients were evaluated.
Results:
The mean osteotomy time was 22.3 ± 3.1 s (range, 17-30 s) and 29.4 ± 3.7 s (range, 25-39 s) in the oscillating saw group and in the new osteotomy instrumentation group, respectively (P < 0.001). The Harris Hip Score (HHS) improved in both groups; the mean HSS was 82.3 ± 2.5 and 83.3 ± 3.5 in the oscillating saw group and new osteotomy instrumentation group at 6 months after surgery, respectively (P = 0.297).
Conclusions:
The original osteotomy instrumentation can be an ideal substitute tool for femoral neck osteotomy in THA, especially when the oscillating saw is unavailable or malfunctioning.