Development of Low-Cost and Portable Pulse Oximeter Device with Improved Accuracy and Accessibility
Hundessa Daba Nemomssa1, Hakkins Raj1
1School of Biomedical Engineering, Jimma Institute of Technology, Jimma University, Jimma, Oromia, Ethiopia.
Purpose:
In a clinical setting, blood oxygen saturation is one of the most important vital sign indicators. A pulse oximeter is a device that measures the blood oxygen saturation and pulse rate of patients with various disorders. However, due to ethical concerns, commercially available pulse oximeters are limited in terms of calibration on critically sick patients, resulting in a significant error rate for measurement in the critical oxygen saturation range. The device's accessibility in developing countries' healthcare settings is also limited due to portability, cost implications, and a lack of recognized need. The purpose of this study was to develop a reliable, low-cost, and portable pulse oximeter device with improved accuracy in the critical oxygen saturation range.
Methods:
The proposed device measures oxygen saturation and heart rate using the reflectance approach. The rechargeable battery and power supply from the smartphone were taken into account, and the calibration in critical oxygen saturation values was performed using Prosim 8 vital sign simulator, and by comparing with a standard pulse oximeter device over fifteen iterations.
Results:
The device's prototype was successfully developed and tested. Oxygen saturation and heart rate readings were both accurate to 97.74% and 97.37%, respectively, compared with the simulator, and an accuracy of 98.54% for the measurement of blood oxygen saturation was obtained compared with the standard device.
Conclusion:
The accuracy of oxygen measurement attained in this study is significant for measuring oxygen saturation for patients in critical care, anesthesia, pre-operative and post-operative surgery, and COVID-19 patients. The advancements made in this research have the potential to increase the accessibility of pulse oximeter in resource limited areas.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse