Original Knee Fixation Device as a Useful Fixation Method during Prostate Intensity-Modulated Radiation Therapy
Noriyuki Kuga1,2, Katsutoshi Shirieda2,3, Yuta Sato2
1Department of Radiological Science, Faculty of Health Sciences, Junshin Gakuen University, Fukuoka, Japan.
A novel knee fixation device (KFD) significantly reduced setup errors in intensity-modulated radiation therapy (IMRT). This improved clinical target volume (CTV) coverage, particularly for prostate cancer patients.
Area of Science:
- Medical Physics
- Radiation Oncology
- Biomedical Engineering
Background:
- Intensity-modulated radiation therapy (IMRT) requires precise patient positioning to ensure accurate radiation delivery.
- Setup errors, including rotational and translational inaccuracies, can compromise treatment efficacy and increase toxicity.
- Existing immobilization techniques may not sufficiently address all sources of setup variability.
Purpose of the Study:
- To evaluate the effectiveness of an original knee fixation device (KFD) in reducing setup errors during IMRT.
- To assess the impact of the KFD on clinical target volume (CTV) coverage.
- To quantify improvements in rotational and translational accuracy with the KFD.
Main Methods:
- A comparative study was conducted with two groups: knee flexion (KF) using KFD (n=16) and knee extension (KE) without KFD (n=15).
- Residual rotational errors (yaw, pitch, roll) and inter-fractional setup errors (A-P, S-I) were measured.
- Inter-fractional margins were calculated, and CTV coverage was evaluated using logistic regression analysis.
Main Results:
- The KF group showed significant improvements in residual rotational errors (yaw, roll) and inter-fractional error in the A-P direction (P < 0.02).
- Inter-fractional margins were reduced by 20.8% in the A-P and 12.6% in the S-I directions in the KF group.
- The odds ratios for CTV coverage were significantly higher in the KF group (2.76 for A-P, 1.74 for S-I).
Conclusions:
- The KFD effectively reduces rotational and translational setup errors in IMRT.
- The use of KFD leads to improved CTV coverage and reduced margins, enhancing treatment precision.
- This original KFD is a potentially valuable tool for improving prostate IMRT immobilization.