Analytical and clinical performance evaluation of a new high-sensitivity cardiac troponin I assay
Shuo Yang1, Qian Zhang1, Boxin Yang1
1Department of Laboratory Medicine, Peking University Third Hospital, Beijing, P.R. China.
Objectives:
To validate the analytical performance and diagnostic accuracy for non-ST-segment elevation myocardial infarction (NSTEMI) with a new high-sensitivity cardiac troponin I (hs-cTnI) assay on the automated light-initiated chemiluminescent assay (LiCA®) platform.
Methods:
Comprehensive analytical validations were performed, and the 99th percentile upper reference limit (URL) from apparently healthy individuals were established. We evaluated the diagnostic performance of the assay for NSTEMI.
Results:
The limit of quantitation (LoQ) were 1.9 ng/L (20% CV) and 5.1 ng/L (10% CV). The sex-specific 99th percentile URLs were 17.6 ng/L (4.2% CV) for men (age 20-79y) and 14.2 ng/L (4.9% CV) for women (age 19-89y) in serum, 14.4 ng/L (4.9% CV) for men (age 19-88y) and 12.9 ng/L (5.2% CV) for women (age 19-87y) in plasma, respectively. Detection rates in healthy individuals were from 98.7 to 99.1%. The correlation coefficient and median bias between LiCA and Architect were 0.985 and 0.1% (-2.0-2.9%) in full analytical range of serum specimens. In lower range (<100 ng/L), LiCA had an overall positive bias 6.7% (-1.6-13.3%), R=0.949. At the specific medical decision levels (15.2, 26.2 and 64.0 ng/L), assay difference was estimated to be <10%. No significant differences on AUC, sensitivity and specificity, NPV and PPV were found between LiCA and Architect for the diagnosis of NSTEMI.
Conclusions:
LiCA hs-cTnI is a precise, highly sensitive and specific assay that meets the requirement of a 3rd generation (level 4) high-sensitivity method. The diagnostic accuracy of LiCA assay for NSTEMI is comparable to the established Architect hs-cTnI assay.
More Related Videos
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome I: Introduction
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Myocarditis II: Clinical Features and Diagnostic Tests