A unified serum IgG4 cut-off level for the diagnosis of IgG4-related disease using a wide array of kits
Songxin Yan1, Yu Peng2, Linlin Cheng1
1Department of Clinical Laboratory, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.
Objectives:
An immunoglobulin G4 (IgG4) level above 1350 mg/L is one of the comprehensive criteria for the diagnosis of IgG4-related disease (IgG4-RD). The purpose of this study was to evaluate the differences in IgG4 levels determined using reagents from two main manufacturers and their concordance with clinical diagnosis.
Methods:
IgG4 levels were measured in 309 patients, including 146, 40, 42, 41, and 40 patients with untreated IgG4-RD, pancreatic cancer, primary Sjogren syndrome, systemic lupus erythematosus, and idiopathic retroperitoneal fibrosis, respectively, and 141 healthy controls. The results obtained using the Binding Site and Siemens reagents were compared in patients with IgG4-RD.
Results:
The serum IgG4 level measured using the Siemens reagent was almost two times that measured using the Binding Site reagent. The rate of IgG4-negative patients, which was 19.9% based on measurement using the Binding Site reagent, was only 4.8% based on measurement using the Siemens reagent (p < .001).
Conclusions:
There were significant differences in serum IgG4 levels based on commonly used reagents from different manufacturers. The IgG4 cut-off level of 1350 mg/L was not suitable for all detection reagents. Clinicians and patients should be cognizant of these differences associated with the specific detection reagents when evaluating the test results.