An Evaluation of the NICCI Monitor in Providing Continuous, Noninvasive Blood Pressure Readings in Children During
Ismail Bekiroglu1, Kwaku Owusu-Bediako1, Julie Rice-Weimer1
1Department of Anesthesiology & Pain Medicine, Nationwide Children's Hospital, Columbus, OH, USA.
Background:
In specific clinical scenarios or with patient comorbid conditions, continuous blood pressure (BP) monitoring may be required as it can be expected to provide superior physiologic monitoring to intermittent techniques. However, continuous BP monitoring requires an arterial cannula (AC) placement, which may be time-consuming, technically challenging, or associated with adverse events. Various noninvasive BP devices have been developed which provide a continuous BP reading. The current study evaluates the accuracy of a novel continuous BP device, the NICCI monitor (NM), in pediatric patients weighing 10 - 40 kg.
Methods:
The study cohort included pediatric patients weighing between 10 and 40 kg, scheduled for surgery for which placement of an AC was planned. Systolic (SBP), diastolic (DBP), and mean arterial (MAP) blood pressure readings were captured from the AC and the NM every second during anesthetic care.
Results:
The study cohort included 24 pediatric patients undergoing major orthopedic, cardiac, and neurosurgical procedures. A total of 146,562 pairs of SBP, DBP, and MAP values from the AC and NC were analyzed. The absolute difference for the NM and the AC's SBP, DBP, and MAP values were 11 ± 9, 10 ± 7, and 10 ± 7 mm Hg, respectively. The difference between the BP values from the NM and the AC was ≤ 10 mm Hg for 57% of the SBP readings, 60% of the DBP readings, and 56% of the MAP readings. The bias was 4, 4, and 6 mm Hg for the SBP, DBP, and MAP, respectively.
Conclusions:
Although there were some technical limitations given patient size, potentially resulting in variation in accuracy, the NM provided a continuous noninvasive BP measurement within clinically useful limits during a significant portion of anesthetic care compared to values obtained from an AC. Ongoing modification of the technology to improve its application in pediatric patients will likely increase its accuracy.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: