Reduced-Dose Full-Body CT in Lymphoma Follow-up: A Pilot Study
Renjun Huang1, Jiulong Yan1, Hongzhi Geng2
1Department of Radiology, the First Affiliated Hospital of Soochow University, Suzhou City, Jiangsu Province 215000, P.R. China.
Background:
How to reduce the radiation dose received from full-body CT scans during the follow-up of lymphoma patients is a concern.
Objective:
The aim of the study was to investigate the image quality and radiation dose of reduced-dose full-body computerized tomography (CT) in lymphoma patients during the follow-up.
Methods:
121 patients were included and divided into conventional CT group (group 1, 120-kVp, n = 61) or reduced-dose CT group (group 2, 100-kVp combined dual-energy CT (DECT), n = 60). 140-kVp polychromatic images and 70-keV monochromatic images were reconstructed from DECT. The abdominal virtual non-enhanced (VNE) images were reconstructed from monochromatic images. Two radiologists rated the overall image quality with a five-point scale and graded the depiction of lesions using a four-point scale. The objective image quality was evaluated using image noise, signal-to-noise ratio, and contrast-to-noise ratio. The radiation dose and image quality were compared between the groups.
Results:
The comparable subjective image quality was observed between 70-keV and 120-kVp images in the neck, while 120-kVp images showed better objective image quality. 70-keV images showed better objective image quality in the chest. While the subjective image quality of abdominal VNE images was inferior to that of true non-enhanced images, the improved objective image quality was observed in VNE images. In the abdominal arterial phase, similar subjective image quality was observed between the groups. Abdominal 70-keV images in the arterial phase showed improved objective image quality. Similar image quality was obtained in the abdominal venous phase between the groups. The effective radiation dose in group 2 showed a significant reduction.
Conclusion:
The application of reduced-dose full-body CT can significantly reduce the radiation dose for lymphoma patients during the follow-up while maintaining or improving the image quality.