Related Experiment Video
Updated: Sep 23, 2025

A Model to Simulate Clinically Relevant Hypoxia in Humans
Published on: December 22, 2016
Relying on pulse oximetry to avoid hypoxaemia and hyperoxia: A multicentre prospective cohort study in patients with
Thierry Boulain1, Mai-Anh Nay1, Pierre-François Dequin2
1Medical-Surgical Intensive Care Unit, Regional Hospital Centre, Orléans, France.
Purpose:
The purpose of this study was to assess the predictive performance of pulse oximetry (SpO2) to rule out hypoxaemia and hyperoxia in critically ill patients.
Methods:
SpO2, arterial oxygenation (SaO2), and arterial partial pressure of oxygen (PaO2) were prospectively and simultaneously measured every 6 h during the first 24 h of intensive care unit admission in a multicentre cohort of critically ill patients suffering acute circulatory failure. Likelihood ratios associated with different cutoff values of SpO2 to rule out hypoxaemia (SaO2 < 90% or PaO2 < 60 mmHg) or hyperoxia (SaO2 > 95% or PaO2 > 100 mmHg) and post-test probabilities were calculated. Mean bias between SpO2 and SaO2 and agreement interval were calculated. Area under the receiver operating characteristics associated with SpO2 to predict different threshold values of SaO2 and PaO2 were calculated.
Results:
Five hundred seventy-one patients (mean [standard deviation] Simplified Acute Physiology Score II: 58.7 [20.1]; mechanically ventilated 75.6%) with 2643 available SaO2 and PaO2 samples and corresponding 2643 SpO2 values were analysed. Mean bias between SpO2 and SaO2 was 1.1%, and its agreement interval ranged from -8.2 to +11.1%. SpO2 cutoff values of 88%, 90%, and 92% left the possibility that 8%-13% of patients had hypoxaemia. SpO2 < 95% left the possibility that 31% of patients had hyperoxia. All calculated areas under the receiver operating characteristics showed a lower limit of their 95% confidence interval below 0.85 CONCLUSION: In this cohort of patients with circulatory failure, SpO2 had poor discriminative ability to rule out hypoxaemia and hyperoxia. Overconfidence upon SpO2 monitoring may be dangerous.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:10Author Spotlight: Exploring Venous Waveforms for Non-Invasive Respiratory Monitoring in Pigs
Published on: March 8, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: