His-bundle pacing to treat an unusual case of chest pain after pacemaker implant
Matteo Parollo1, Giulio Zucchelli1, Mario Giannotti Santoro1
1Second Department of Cardiology, Cardiac Thoracic and Vascular Department - Pisa University Hospital, Pisa, Italy.
Insights
Painful right ventricular pacing, often linked to dyssynchrony, can be effectively treated with His-bundle pacing (HBP). This innovative approach offers a pain-free solution for patients experiencing discomfort during standard pacing.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- A 63-year-old male with coronary artery disease and prior bypass surgery presented with chest pain.
- He had a pacemaker implanted 5 months prior for atrioventricular block, with normal device parameters post-implantation.
Observation:
- Chest pain was exclusively triggered by right ventricular (RV) pacing, despite normal lead function and no evidence of perforation.
- Electrophysiology study confirmed painful RV pacing, suggesting pacing-induced dyssynchrony as the potential cause.
Findings:
- His-bundle pacing (HBP) was successfully implemented using a specialized catheter and cardiac resynchronization therapy generator.
- The patient experienced complete pain relief during HBP and remained asymptomatic at 6-month follow-up, with normal device function.
Implications:
- Painful RV pacing is a rare, potentially underdiagnosed condition possibly caused by ventricular dyssynchrony.
- His-bundle pacing offers a novel and effective therapeutic strategy, achieving physiological activation and resolving pacing-induced pain.
Abstract:
A 63-year-old man with hypertension and 3-vessel coronary artery disease previously treated with coronary artery bypass graft was admitted to our emergency room complaining of chest pain. He had undergone pacemaker implant 5 months before due to paroxysmal advanced atrioventricular block. Electrocardiography and troponin testing were unremarkable. Echocardiography and chest X-ray ruled out lead displacement and perforation. Interrogation showed normal parameters [right atrium: impedance 550 Ohm bipolar, sensing 2.4 mV bipolar; threshold 0.50 V/0.4 ms bipolar; right ventricle (RV): impedance 580 Ohm bipolar, sensing > 25 mV bipolar; threshold 1.5 V/0.4 ms bipolar and 0.4 V/0.4 ms unipolar]. Pain was evoked only during RV pacing. An electrophysiology study demonstrated painful RV pacing from multiple sites. We hypothesized that pain was associated with pacing-induced dyssynchrony. His-bundle pacing (HBP) was considered as a solution. We achieved HBP with a bipolar fixed-screw catheter connected to a cardiac resynchronization therapy pacemaker generator. During HBP above threshold (4.00 V/1.00 ms) the patient did not complain of any pain. He was discharged 3 days later pain-free with His-bundle lead amplitude set at 5.00 V/1.00 ms. After 6 months the patient was asymptomatic, with the device showing normal functioning. This is the first clinical experience of painful RV pacing treated with HBP. <Learning objective: Painful right ventricular pacing in the absence of perforation is a rare but potentially underdiagnosed condition. Ventricular dyssynchrony could represent the underlying mechanism. Physiological electromechanical activation achieved via His-bundle pacing could represent an effective therapeutic option.>.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Dysrhythmias VI: Management of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias