Related Experiment Video
Updated: Sep 22, 2025

Endoscopic Ultrasound-Guided Biliary Drainage: Endoscopic Ultrasound-Guided Hepaticogastrostomy in Malignant Biliary Obstruction
Published on: March 25, 2022
A novel bipolar electrocautery-enhanced delivery system with a lumen-apposing metal stent for EUS-guided drainage: A
Seok Jung Jo1, Jong Ho Moon1, Yun Nah Lee1
1Department of Internal Medicine, Digestive Disease Center and Research Institute, SoonChunHyang University School of Medicine, Bucheon, Korea.
Background:
An electrocautery-enhanced delivery system with a lumen-apposing metal stent (LAMS) is available for one-step endoscopic ultrasound-guided transmural drainage (EUS-TD). Bipolar electrosurgery has several potential clinical advantages, including reduced collateral thermal damage, enhanced hemostasis, and no requirement for a return electrode plate. In this study, we compared the technical feasibility and safety of a newly developed bipolar electrocautery-enhanced delivery system with a conventional delivery system for EUS-TD using a LAMS in a porcine model.
Method:
Ten days before the study, 12 mini pigs underwent common bile duct ligation for EUS-guided gallbladder drainage. Transenteric puncture was performed, followed by placement of a guidewire. In six pigs, a bipolar electrocautery-enhanced delivery system with LAMS (Hot SPAXUS) was inserted over the guidewire and advanced into the gallbladder, without prior dilation of the tract, by applying a bipolar cut current. In the remaining six pigs, a conventional delivery system with LAMS (Cold SPAXUS) was inserted after tract dilatation using a cystotome. The stent was removed after 4 weeks.
Results:
In all pigs, the stent was successfully inserted and deployed in the gallbladder without adverse events. The mean procedure time of EUS-TD was significantly lower in the Hot SPAXUS group than that of the Cold SPAXUS group (mean ± standard deviation: 188.7 ± 5.2 vs 449.5 ± 97.5 s, P = .0019). Stent migration was not observed, and all stents were removed successfully.
Conclusions:
Endoscopic ultrasound-guided transmural drainage using a bipolar electrocautery-enhanced LAMS is feasible for reducing the procedure time while maintaining the high success rate and safety of conventional LAMS.
More Related Videos
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020