ARISE: First-In-Human Evaluation of a Novel Stent Graft to Treat Ascending Aortic Dissection

Eric E Roselli1, Marvin D Atkins2, William Brinkman3

  • 1Department of Thoracic and Cardiovascular Surgery, Aortic Center, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH, USA.

Insights

The ARISE study shows promising early results for a novel stent graft in treating high-risk ascending aortic dissection patients. This investigational device offers a potential new option for those unsuitable for traditional surgery.

Area of Science:

  • Cardiovascular Surgery
  • Medical Devices
  • Interventional Cardiology

Background:

  • Ascending aortic dissection carries a 10-20% operative mortality rate.
  • A significant portion of patients are ineligible for open surgical repair.
  • The ARISE study investigates a novel GORE® Ascending Stent Graft (ASG).

Purpose of the Study:

  • To assess the early feasibility of using the investigational GORE® Ascending Stent Graft (ASG).
  • To treat patients with ascending aortic dissection.

Main Methods:

  • Prospective, multicenter, non-randomized, single-arm study.
  • Enrollment of high-surgical-risk patients with suitable anatomy.
  • Transfemoral delivery of the ASG under fluoroscopic guidance.

Main Results:

  • Nineteen patients (mean age 75.7 years) with acute DeBakey type I or II dissection were treated.
  • Successful device delivery and deployment in all patients.
  • 30-day MACCE: 15.8% mortality, 5.3% disabling stroke, 5.3% myocardial infarction.

Conclusions:

  • The ARISE early feasibility study demonstrates promising results for the GORE® Ascending Stent Graft.
  • The device offers a potential new treatment avenue for ascending aortic dissection.
Abstract

Related Concept Videos