Comparison of Doppler Flow Velocity and Thermodilution Derived Indexes of Coronary Physiology

Ozan M Demir1, Coen K M Boerhout2, Guus A de Waard3

  • 1British Heart Foundation Centre of Excellence and National Institute for Health Research Biomedical Research Centre at the School of Cardiovascular Medicine and Sciences, King's College London, London, United Kingdom.

Insights

Coronary flow reserve (CFR) measured by thermodilution overestimates Doppler-derived CFR. A thermodilution threshold of 2.5 provides better accuracy for diagnosing reduced coronary flow reserve than the commonly used 2.0 threshold.

Area of Science:

  • Cardiology
  • Medical Imaging
  • Hemodynamics

Background:

  • Coronary flow reserve (CFR) is crucial for assessing myocardial perfusion.
  • Doppler-derived CFR measurements are common, but their correlation with thermodilution is unclear.
  • Diagnostic thresholds for CFR require validation across different measurement techniques.

Purpose of the Study:

  • To compare Doppler flow velocity and thermodilution-derived indexes for CFR measurement.
  • To establish optimal thermodilution-based diagnostic thresholds for CFR.
  • To evaluate the correlation between Doppler and thermodilution-derived microvascular resistance.

Main Methods:

  • An international multicenter registry collected data from patients undergoing both Doppler and thermodilution CFR measurements.
  • Physiological data from 250 vessels in 149 patients were analyzed.
  • Correlation and agreement analyses were performed, including Bland-Altman analysis.

Main Results:

  • A modest correlation (r²=0.36) was found between thermodilution-derived CFR (CFRthermo) and Doppler-derived CFR (CFRDoppler).
  • CFRthermo overestimated CFRDoppler (mean bias 0.59).
  • A CFRthermo threshold of 2.5 showed better sensitivity (75.5%) and specificity (81.3%) than the common threshold of 2.0 for predicting CFRDoppler < 2.5. Microvascular resistance indexes showed weak correlation.

Conclusions:

  • Thermodilution-derived CFR overestimates Doppler-derived CFR, with only a modest correlation between the two methods.
  • The standard thermodilution CFR threshold of 2.0 has low sensitivity for detecting reduced CFR.
  • A thermodilution threshold of 2.5 offers improved diagnostic accuracy for assessing coronary flow reserve.
Abstract