Two-dimensional shear wave elastography with two different systems for the diagnosis of breast lesions
Ji-Feng Yu1,2,3, Shen Zhang1,2,3, Hao-Hao Yin2,3
1Department of Medical Ultrasound, Shanghai Tenth Hospital, School of Clinical Medicine of Nanjing Medical University, Shanghai, P.R. China.
Background:
Two-dimensional (2D) - shear wave elastography (SWE) has made promising advances in the diagnostic of breast lesions. However, few studies have assessed whether the diagnostic effectiveness of different platforms employing 2D-SWE is equal or different.
Objective:
To compare the diagnostic effectiveness of 2D-SWE techniques from two different systems in differentiating malignant breast lesions from benign ones.
Methods:
A total of 84 breast lesions were retrospectively analyzed by experienced radiologists using 2D-SWE on two ultrasound systems, i.e. system-1 (LOGIQ E9 system, GE Healthcare, Wauwatosa, WI, USA), and system-2 (Aixplorer US system, SuperSonic Imagine, Aix-en-Provence, France). Qualitative and quantitative parameters including color sign, the maximum elasticity modulus values (E-max), the mean elasticity modulus values (E-mean) and standard deviation (E-sd) of elasticity modulus values in two 2D-SWE systems were analyzed. The diagnostic performance between system-1 and system-2 were evaluated in terms of the areas under the receiver operating characteristic curves (AUROCs).
Results:
Among the 84 lesions in this study, 66 (78.6%) were benign and 18 (21.4%) were malignant. E-max in system-1 showed the best diagnostic performance with a cut-off value of 174.5 kPa with the associated sensitivity and specificity of 100.0% and 80.3% respectively. Meanwhile, E-sd in system-2 displayed the best diagnostic performance with a cut-off value of 12.7 kPa, with the associated sensitivity and specificity of 94.4% and 80.3% respectively. The diagnostic performance of the two 2D-SWE systems was not statistically different according to receiver operating characteristic curve (ROC) analysis of E-max, E-mean, and E-sd.
Conclusion:
For identifying breast lesions, system-1 and system-2 appear to be similar in diagnostic performance. However, different cut-off values for different parameters might be selected to obtain the best diagnostic performance for the two 2D-SWE systems.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies II: Ultrasonography
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Ultrasonography
During an ultrasonography procedure, a handheld device called...