Evaluating image quality and optimal parameters for non-linear blending dual-energy computed tomography images of
Liying Lin1, Li Han2,3,4, Shaowei Jia5
1First Central Clinical College, Tianjin Medical University, Tianjin, China.
Background:
Blending technology is usually used to improve quality of dual-energy computed (DECT) images.
Objectives:
To evaluate the blended DECT image qualities by employing the Blending-Property-Map (BP-Map) and elucidating the optimal parameters with the highest signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR).
Methods:
Sixty pairs of 80 kV and 140 kV CT images are blended non-linearly by four methods. Protocol A uses the fixed values of blending width (BW) and blending center (BC); Protocol B uses the values of BW = (CThepatic portal vein - CThepatic parenchymal) / 2 and BC = (CThepatic portal vein + CThepatic parenchymal) / 2; Protocol C uses a BW ranging from 10 to 100 HU at an interval of 10 HU and BC = (CThepatic portal vein + CThepatic parenchymal) / 2; Protocol D uses the BP-Map that covers all possible values of BW and BC.
Results:
When using CT value of adipose tissue as noise, the calculated SNR and CNR of optimal blending width and blending center were 123.22±41.73 and 9.00±3.52, respectively, by the BP-Map in the protocol D. By employing the CT value of back muscle as noise, the SNR and CNR of the best-blended images were 75.90±14.52 and 6.39±2.37, respectively. The subjective score of protocol D was 4.88±0.12.
Conclusions:
Compared to traditional blending methods, the BP-Map technique can determine the optimal blending parameter and provide the best-blended images with the highest SNR and CNR.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:20Contrast-Enhanced Subharmonic Aided Pressure Estimation SHAPE Using Ultrasound Imaging with a Focus on Identifying Portal Hypertension
Published on: December 5, 2020
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Hepatic Portal System
At its core, the hepatic portal vein is the result of a confluence of the superior and inferior mesenteric veins along with the splenic vein. Each of these veins has a unique role. The superior mesenteric vein is...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: Kidney, Ureter, and Bladder Studies