Performance evaluation of the Elecsys PIVKA-II and Elecsys AFP assays for hepatocellular carcinoma diagnosis
Henry L Y Chan1, Arndt Vogel2, Thomas Berg3
1Faculty of Medicine The Chinese University of Hong Kong Hong Kong China.
Background And Aims:
Prothrombin induced by vitamin K absence-II (PIVKA-II) is a serum biomarker linked to hepatocellular carcinoma (HCC), showing superiority to alpha-fetoprotein (AFP) for early disease detection. We aimed to assess the clinical and analytical performance of the Elecsys® PIVKA-II immunoassay in diagnosing HCC and evaluate PIVKA-II's technical performance.
Methods:
Serum samples from adult cases (i.e. patients with a first-time HCC diagnosis; n = 168) and disease controls (i.e. patients without HCC with an at-risk condition; n = 208) were assessed. An AFP cut-off of 20 ng/mL was used to differentiate between HCC cases and disease controls. Clinical performance of the Elecsys PIVKA-II assay was compared with that of comparator assays (Lumipulse G PIVKA-II, μTASWako DCP, ARCHITECT PIVKA-II) using receiver operating characteristic curve analysis to determine the area under the curve (AUC) values.
Results:
The Elecsys PIVKA-II assay compared favorably with comparator assays. Using a 28.4 ng/mL cut-off, the Elecsys PIVKA-II assay detected HCC with 86.9% sensitivity and 83.7% specificity. Clinical performance of the Elecsys PIVKA-II assay (AUC: 90.8%) was equivalent to that of comparator assays (AUC: 88.3-89.6%). Relatively high PIVKA-II concentrations were observed for cholangiocarcinoma and pancreatic cancer with the Elecsys assay in specificity panel analyses, indicating that high PIVKA-II concentrations should not be used alone in the absence of other clinical data.
Conclusions:
The Elecsys PIVKA-II assay showed good analytical performance under routine laboratory conditions, comparing favorably with comparator assays. These findings support the suitability of the Elecsys PIVKA-II assay as an aid in HCC diagnosis.