Use of a Viral Filter to Reduce Exposure to Exhaled Aerosol Does Not Affect Methacholine Dose Delivery During
Yosuf W Subat1, Todd J Meyer2, Keith D Torgerud3
1Division of Pulmonary, Critical Care, and Sleep Medicine, Department of Medicine, Mayo Clinic, Rochester, Minnesota.
Background:
Methacholine challenge testing (MCT) is a common bronchoprovocation technique used to assess airway hyper-responsiveness. We previously demonstrated that the addition of a viral filter to the nebulizer exhalation limb substantially reduced expelled particles during MCT. Our aim was to evaluate whether this modification affects the delivered dose of methacholine.
Methods:
A mechanical ventilator was connected to a lung simulator with breathing frequency 15 breaths/min, tidal volume 500 mL, inspiratory-expiratory ratio 1:1, with a sinusoidal waveform. We compared methacholine dose delivery using the Hudson Micro Mist or AeroEclipse II BAN nebulizers powered by either a dry gas source or a compressor system. A filter placed in line between the nebulizer and test lung was weighed before and after 1 min of nebulized methacholine delivery. Mean inhaled mass was measured with and without a viral filter on the exhalation limb. Dose delivery was calculated by multiplying the mean inhaled mass by the respirable fraction (particles < 5 μm) and inhalation time. Unpaired t test was used to compare methacholine dose delivery with and without viral filter placement.
Results:
The addition of a viral filter did not significantly affect methacholine dose delivery across all devices tested. Using a 50-psi dry gas source, dose delivered with or without a viral filter did not differ with the Hudson (422.3 μg vs 282.0 μg, P = .11) or the AeroEclipse nebulizer (563.0 μg vs 657.6 μg, P = .59). Using the compressor, dose delivered with and without a viral filter did not differ with the Hudson (974.0 μg vs 868.0 μg, P = .03) or the AeroEclipse nebulizer (818.0 μg vs 628.5 μg, P = .42).
Conclusions:
The addition of a viral filter to the nebulizer exhalation limb did not affect methacholine dose during bronchoprovocation testing. Routine use of a viral filter should be considered to improve pulmonary function technician safety and infection control measures during the ongoing COVID-19 pandemic.
Related Concept Videos
Inhaled Medications
COPD: Management Using Bronchodilators and Corticosteroids
Antiasthma Drugs: Methylxanthines
Theophylline is thought to inhibit phosphodiesterase enzymes, increasing intracellular levels of cyclic adenosine monophosphate (cAMP) and cyclic guanosine monophosphate (cGMP). This rise in cAMP and cGMP concentrations stimulates cardiac function,...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...