Oscillation Transmission of Modern High-Frequency Neonatal Ventilators Under Different Lung Mechanics Conditions
Roberta Centorrino1, Valentina Dell'Orto1, Charlotte Boussard1
1Division of Pediatrics and Neonatal Critical Care, "A.Beclere" Medical Center, Paris Saclay University Hospitals, APHP, Paris, France.
Background:
High-frequency oscillatory ventilation (HFOV) is widely used in neonatal critical care, and several modern ventilators using different technologies are available to provide HFOV. These devices have different technical characteristics that might interact with patient lung mechanics to influence the effectiveness of ventilation. To verify this, we studied the oscillation transmission of 5 neonatal oscillators in a lung model mimicking the mechanical patterns commonly observed in neonatal practice.
Methods:
This was a benchtop, in vitro, physiological, pragmatic study using a model mimicking airways and lung of a 1-kg preterm neonate and the following patterns: normal (compliance: 1.0 mL/cm H2O, resistance: 50 cm H2O/L/s), restrictive (compliance: 0.3 mL/cm H2O, resistance: 50 cm H2O/L/s), and mixed mechanics (compliance: 0.3 mL/cm H2O, resistance: 250 cm H2O/L/s). Several permutations of HFOV parameters (variable mean airway pressure or amplitude or frequency protocols) were tested. Oscillations were measured with a dedicated pressure transducer validated for use during HFOV, and oscillatory pressure ratio (OPR) was calculated to estimate the oscillation transmission.
Results:
Overall OPR (calculated on all experiments) was significantly different between ventilators and the mechanical patterns (both P < .001). Different ventilators and patterns accounted for 35.6% and 20.6% of the variation in oscillation transmission, respectively. Sub-analyses per changing amplitude or frequency protocols and multivariate regressions showed that VN500 (standardized β coefficient [St.β]: 0.548, P < .001) and Fabian HFO (St.β: 0.421, P < .001; adjusted R2: 0.615) provided the best oscillation transmission. Fabian HFO also delivered oscillations with the lowest variability when increasing amplitude.
Conclusions:
In an experimental setting mimicking typical neonatal lung disorders, the oscillation transmission was more dependent on the ventilator model than on the mechanical lung conditions at equal HFOV parameters. Fabian HFO and VN500 provided better oscillation transmission overall, and when increasing amplitude, Fabian HFO delivered oscillations with the lowest variability.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...