Related Experiment Video
Updated: Sep 22, 2025

Author Spotlight: Implications of Non-Nutritive Sucking on Speech Emergence and Infant Development
Published on: April 19, 2024
A Modified Method for Measuring Pressure Drop in Non-medical Face Masks with Automated Data Acquisition and Analysis
Alexander Herman1, Daniel Porter1, Dana Rottach1
1Center for Devices and Radiological Health, US Food and Drug Administration, 10903 New Hampshire Ave., Building 62, Room 2229, Silver Spring, MD 20993.
Background:
Non-medical face masks, such as face coverings donned by the general population play an important role in reducing transmission of respiratory pathogens. Pressure drop or breathability of such masks is an important attribute especially with the advent of new standards such as ASTM F3502-21 that have specified pressure drop limits for general use of face coverings. Although several standards are available that discuss pressure drop measurement techniques, the methodologies reported are typically complex or are part of more sophisticated and expensive instruments. Thus, the applicability of such methods is often limited to medical device manufacturers.
Objective And Methods:
This manuscript adapts from the pressure drop measurements proposed in British Standard EN 14683:2019 and describes a methodology to create a simple 3D printed model of a pressure rig for measuring the breathing resistance across non-medical face masks. The method also enables real time pressure drop data acquisition and analysis of multiple samples or batches using Python and MATLAB scripts.
Results:
We performed a validation study by comparing the pressure drop obtained for one brand of respirators with our set up and compared it with data obtained by traditional means by CDC. An unpaired two-tailed student t-test (n=3) between the two means implied no statistically significant difference.
Conclusion:
The method we have developed can be easily implemented at community levels for characterizing the breathability of non-medical grade face masks.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: