Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Paced P-wave duration as a significant predictor for atrial high-rate episodes in patients with cardiac implantable
José M Rubio Campal1, Alberto Sánchez Arjona1, Camila S García Talavera1
1Arrhythmia Unit, Cardiology Department, Fundación Jiménez Díaz - Quirón Salud, Madrid, Spain.
Background:
Atrial pacing (AP) can unmask or aggravate a preexisting interatrial block (IAB). The aim of our study was to determine whether AP is associated with the development of atrial high-rate episodes (AHRE) during follow-up.
Methods:
Patients with dual-chamber cardiac implantable electronic devices (CIEDs), no previous documented atrial fibrillation, and with a 6-month minimum follow-up were included. In all patients, sinus and paced P-wave duration were measured. AHRE was defined as an episode of atrial rate ≥225 bpm with a minimum duration of 5 min, excluding those documented during the first 3 months after implantation.
Results:
A total of 220 patients were included (75 ± 10 years, 61% male). After a mean follow-up of 59 ± 25 months, 46% of patients presented AHRE. Mean paced P-wave duration was significantly longer than the sinus P-wave duration (154 ± 27 vs. 115 ± 18 ms; p < .001). Sinus and paced P-waves were significantly longer in those who developed AHRE (sinus: 119 ± 20 vs. 112 ± 16; p = .006; paced: 161 ± 29 vs. 148 ± 23; p < .001). A paced P-wave ≥160 ms was the best predictor of AHRE, especially those lasting >24 h (odds ratio [OR] 4.2 [95% confidence interval (CI)] [1.6-11.4]; p = .004).
Conclusions:
AP significantly prolongs P-wave duration and is associated with further development of AHRE. A paced P-wave ≥160 ms is a strong predictor of AHRE and should be taken into consideration as a new definition of IAB in the presence of AP.
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...