Pulse-Echo Technique to Compensate for Laminate Membrane Transmission Loss in Phantom-Based Ultrasonic Attenuation
Karthik Nagabhushana1, Qiuyu Wang2, Aiguo Han1
1Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering, and National Center for Supercomputing Applications, University of Illinois Urbana-Champaign, Urbana, IL, USA.
Objectives:
Accurately measuring the attenuation coefficient (AC) of reference phantoms is critical in clinical applications of quantitative ultrasound. Phantom AC measurement requires proper compensation of membrane transmission loss. Conventional methods require separate membrane samples to obtain membrane transmission loss. Unfortunately, separate membrane samples are often unavailable. A pulse-echo approach is proposed herein to compensate for membrane transmission loss without requiring separate membrane samples.
Methods:
The proposed method consists of the following steps. First, the insertion loss, caused by phantom attenuation and membrane transmission loss, is measured. Second, the membrane reflection coefficient is measured. Third, the unknown acoustic parameters of the membrane and phantom material are estimated by fitting theoretical reflection coefficient to the measured one. Finally, the fitted parameters are used to estimate membrane transmission loss and phantom AC. The proposed method was validated through k-Wave simulations and phantom experiments. Experimental AC measurements were repeated on 5 distinct phantoms by 2 operators to assess the repeatability and reproducibility of the proposed method. Five transducers were used to cover a broad bandwidth (0.7-16 MHz).
Results:
The acquired AC in the simulations had a maximum error of 0.06 dB/cm-MHz for simulated phantom AC values ranging from 0.5 to 1 dB/cm-MHz. The acquired AC in the experiments had a maximum error of 0.045 dB/cm-MHz for phantom AC values ranging from 0.28 to 1.48 dB/cm-MHz. Good repeatability and cross-operator reproducibility were observed with a mean coefficient of variation below 0.054.
Conclusion:
The proposed method simplifies phantom AC measurement while providing satisfactory accuracy and precision.