Related Experiment Video
Updated: Sep 21, 2025

The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
Which Safe Zone Is Safe in Total Hip Arthroplasty? The Effect of Bony Impingement
Markus Weber1,2, Frederik von Kunow2, Moritz Innmann3
1Faculty of Medicine, University of Regensburg, 93053 Regensburg, Germany.
Abstract:
"Safe zones" for cup position are currently being investigated in total hip arthroplasty (THA). This study aimed to evaluate the impact of bony impingement on the safe zone and provide recommendations for cup position in THA. CT scans were performed on 123 patients who underwent a cementless THA. Using the implant data and bone morphology from the CT scans, an impingement detection algorithm simulating range of motion (ROM) determined the presence of prosthetic and/or bony impingement. An impingement-free zone of motion was determined for each patient. These zones were then compared across all patients to establish an optimized impingement-free "safe zone". Bony impingement reduced the impingement-free zone of motion in 49.6% (61/123) of patients. A mean reduction of 23.4% in safe zone size was observed in relation to periprosthetic impingement. The superposition of the safe zones showed the highest probability of impingement-free ROM with cup position angles within 40-50° of inclination and 20-30° of anteversion in relation to the applied cup and stem design of this study. Virtual ROM simulations identified bony impingement at the anterosuperior acetabular rim for internal rotation at 90° of flexion and at the posteroinferior rim for adduction as the main reasons for bony impingement.