Summary
Pedestal high-contrast gratings (PHCG) offer enhanced label-free biosensing performance. These PHCGs demonstrate superior sensitivity and lower detection limits for biomarkers compared to conventional HCGs.
Area of Science:
- Photonics
- Nanotechnology
- Biomedical Engineering
Background:
- High-contrast gratings (HCG) are effective for label-free biomarker detection due to sharp optical resonances.
- Conventional HCG designs have limitations in sensitivity and detection capabilities.
Purpose of the Study:
- To experimentally evaluate the performance of pedestal HCG (PHCG) for biosensing.
- To compare the sensitivity and detection limits of PHCG with conventional HCG.
Main Methods:
- Fabrication and characterization of pedestal HCG (PHCG) and conventional HCG.
- Measurement of bulk refractive index sensitivity and surface sensitivity using model analyte layers (Al2O3, HfO2, TiO2).
- Avidin-biotin binding assay in solution to determine limit of detection (LoD) and limit of quantification (LoQ).
Main Results:
- PHCG exhibited an 11.2% improvement in bulk refractive index sensitivity (536 nm/RIU vs. 482 nm/RIU).
- PHCG showed a 10.5% improvement in surface sensitivity.
- PHCG achieved a lower limit of detection (LoD) of 2.1 ng/mL and limit of quantification (LoQ) of 85 ng/mL for avidin detection, compared to conventional HCG (LoD: 3.2 ng/mL, LoQ: 213 ng/mL).
Conclusions:
- Pedestal HCG demonstrates significantly enhanced performance for label-free biosensing.
- The improved sensitivity and lower detection limits of PHCG make it highly suitable for detecting low analyte concentrations.
- PHCG holds great potential for advanced biosensing applications.
Keywords:
atomic layer depositionavidinbiosensingbiotinbulk refractive index sensitivityhigh-contrast gratingnanofabricationrefractometric sensingsilicon nanostructuressurface sensitivity