Inter-rater reliability of pressure biofeedback unit among individuals with and without chronic low back pain
Muhammad Khan1, Hamayun Zafar2, Syed Amir Gilani3
1Dr. Muhammad Khan PT, MSPT (UK), Institute of Physical Therapy & Rehabilitation Jinnah Sindh Medical University, Karachi, Pakistan.
Objectives:
To determine the inter-rater reliability of pressure biofeedback unit among individuals with and without chronic low back pain.
Methods:
This cross-sectional survey was conducted from February 2021 to March 2021 at the Physiotherapy Department of the Sindh Institute of Physical Medicine and Rehabilitation. Sixteen subjects which were recruited with and without chronic low back pain.(CLBP). During the test, abdominal drawing in movement was monitored by measuring a change in pressure detected in PBU. Each test was performed once by two trained assessors with period of seven days. Reliability indices of Pressure Biofeedback (PBU) measures including the Intraclass correlation coefficient [ICC] and Band Altman plot were analyzed.
Results:
This study found an intra-class correlation coefficient (ICC) = 0.94 [95% confidence interval (CI) (0.37, 0.99] for inter-rater reliability in healthy individuals, and an ICC of 0.97 (95% CI 0.97, 0.98) for inter-rater reliability in CLBP. The interrater agreement (Limits of Agreement-LOA=5.92, -3.9 mmHg) in CLBP and the interrater agreement (LOA=5.75, -3.25 mmHg) in healthy individuals were within the limits of agreement on 95% of occasions.
Conclusion:
Pressure Biofeedback Unit has showed excellent inter-rater reliability in measuring Transverse Abdominis muscle activity for individuals with and without chronic LBP.
More Related Videos
06:28Biomechanical Changes Related to Low Back Pain: An Innovative Tool for Movement Pattern Assessment and Treatment Evaluation in Rehabilitation
Published on: December 13, 2024
05:10Back Mechanical Sensitivity Assessment in the Rat for Mechanistic Investigation of Chronic Back Pain
Published on: August 30, 2022
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...