Low cost circulatory pressure acquisition and fluid infusion rate measurement system for clinical research
Rachel Smith1, Amelia Rolfe1, Chris Cameron1
1Department of Mechanical Engineering, University of Canterbury, New Zealand.
Abstract:
Acquiring patient physiological waveforms is useful for studying hemodynamic management and developing medical monitoring systems. A low cost, Arduino controlled data acquisition system acquires arterial pressure waveforms (Edwards Lifesciences TruWave compatible) and measures fluid infusion rate using hanging scales. This system can be used at the same time as a clinical monitor, enabling recording of patient arterial pressure and fluid delivery for clinical research. The system is powered via a USB connection, which additionally provides serial output, aiding compatibility and customisation. A simple software user interface, developed in Python, shows outputs. Each data acquisition system, including all necessary connection cables costs ~US$90 and is multiple-use.
More Related Videos
08:51Intra-cardiac Side-Firing Light Catheter for Monitoring Cellular Metabolism using Transmural Absorbance Spectroscopy of Perfused Mammalian Hearts
Published on: May 12, 2019
07:55A Multi-Parametric Islet Perifusion System within a Microfluidic Perifusion Device
Published on: January 26, 2010
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: